When the lower display is in the frequency or pulse mode, the frequency
level menu is added after the auto output main menu. The options available
in this menu are [FREQ LEVEL], [NEXT], and [DONE]. The [FREQ LEVEL]
option is used to adjust the amplitude of the wave. [NEXT] is used to access
the contrast main menu, and [DONE] returns to the home menu.
When the calibrator is in RTD CUSTOM mode, the RTD custom setup menu,
is inserted after the main menu. Options [SET CUSTOM], [NEXT], and
[DONE] are available. [SET CUSTOM] is used to enter a custom PRT into
the calibrator. Refer to Section 4.1-8a for instructions. [NEXT] is used to
enter the contrast main menu, and [DONE] to return to the home menu.
The pressure zeroing main menu is the final variation to choosing [MORE] in
the main menu. It has the options [ZERO ], used to zero pressure,
[NEXT] and [DONE], which have the same function as above. Refer to the
Section 5.3 for instructions on zeroing.
The parameter selection menu is called up when [UPPER] or [LOWER] is
selected from the main menu. It contains the following options: [SELECT],
[NEXT], and [DONE]. When the display is selected, a parameter will start to
flash. Use the [SELECT] option to change the parameter, and the [NEXT]
option to switch to another variable. [DONE] returns to the home menu and
enables the selected mode.