EasyManua.ls Logo

Ametek JOFRA ASC300 - Measuring Pressure

Ametek JOFRA ASC300
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Select [mA LOOP] as primary parameter in the upper display.
2. Connect the calibrator to transmitter current loop terminals, as shown in
Figure 17.
Figure 17. Connection Using Current Loop
5.3 Measuring Pressure
Note: Pressure is not read from modules with frequency or pulse train mode
enabled.
Warning!
To avoid a violent release of pressure in a pressurized system, shut off the
valve and slowly bleed off the pressure before you attach the pressure
module to the pressure line.
Caution
To avoid mechanically damaging the pressure module, never apply more
than 13.5 Nm/10 ftlbs. of torque between the pressure module fittings, or
between the fittings an the body of the module.
To avoid damaging the pressure module from overpressure, never apply
pressure above the rated maximum printed on the module.
To avoid damaging the pressure module from corrosion, use it only with
specified materials. Refer to the pressure module documentation for material
compatibility.
23
!
!