29
3. To switch back to local operation enter LOCAL at the terminal. This
command also turns off LOCKOUT if it was on. For more information on
commands refer to the Remote Commands section.
7.3 Using Commands
7.3-1 Command types
Refer to the Section on Remote Commands for all available commands.
The calibrator may be controlled using commands and queries. All
commands may be entered using upper or lower case. The commands are
divided into the following categories:
Calibrator Commands
Only the calibrator uses these commands. For example
LOWER_MEAS DCV
tells the calibrator to measure voltage on the lower display.
Common Commands
Standard commands used by most devices. These commands always begin
with an "*". For example
*IDN?
tells the calibrator to return its identification.
Query Commands
Commands that ask for information. They always end with a "?". For
example:
FUNC?
Returns the current modes of the upper and lower displays.
Compound Commands
Commands that contain more than one command on one line. For
example:
LOWER_MEAS RTD; RTD_TYPE CU10
Sets the calibrator to measure RTD in the lower display and sets RTD
type to Cu 10.
Overlapped Commands
Commands that require more time to execute than normal. The command
*WAI can be used after the overlapped command to tell the calibrator to wait