Figure 19. Connections for Testing an Output Device
6.2 Calibrating an I/P Device
The following steps show how to calibrate a device that controls pressure:
1. Select upper display from the Main Menu, and select pressure from the
primary parameters. 
2. Switch to lower display from the Main Menu, and select current source
[mA out] from the primary parameters.
3. Connect the calibrator to the device as shown in Figure 20. The
calibrator will simulate the transmitter current and measure the output
pressure.
4. Enter a current using the keypad.
Figure 20. Calibrating an I/P Device
25