To measure pressure, follow these steps:
1. Connect the pressure module to the calibrator as shown in Figure 18.
The calibrator can measure pressure on both the upper and the lower
display. This makes it possible to measure pressure in two different
units at the same time.
2. Switch to either upper or lower display from the Main Menu.
3. Select [PRESSURE] from the primary parameters.
4. Select the desired measuring unit.
5. Zero the pressure module. The zero function on the calibrator can be
found in the pressure zeroing menu. For zeroing absolute modules see
3.4.1.
Figure 18. Measuring Pressure Transmitter
6. Using the Upper and the Lower Display for
Calibration and Testing
6.1 Testing an Input or Indicating Device
To test and calibrate actuators, recording, and indicating devices using the
source functions, follow these steps:
1. Select the lower display and choose the correct primary parameter.
2. Switch to [OUT] in the input/output control.
3. Connect the leads to the instrument and the calibrator as shown in
Figure 19.