4.8 Sourcing Thermocouples
To source a thermocouple use the following steps:
1. Connect the thermocouple leads to the appropriate polarized TC
miniplug, and insert the plug into the TC terminals on the calibrator, as
shown in Figure 13.
2. Select lower display from the Main Menu, and choose thermocouple
[TC] from the primary parameters.
3. Choose output [OUT] from the input/output control.
4. Select the desired thermocouple type from the sensor types.
5. Enter the temperature using the keypad.
Figure 14. Connections for Outputting RTDs
4.9 Sourcing Ohms/RTDs
To source an RTD, follow these steps:
1. Select lower display from the Main Menu, and choose [RTD] from the
primary parameters.
2. Choose output [OUT] from the input/output control, and select RTD
type from the sensor types.
3. Connect the calibrator to the instrument being tested, as in Figure 14.
4. Enter the temperature or resistance using the keypad.
20