FREQ_UNIT
Sets the unit for frequency. There are three ranges of frequencies for
frequency and pulse modes, CPM (cycles per minute), Hz, and kHz. Use
this command to select the right range. For example:
FREQ_UNIT HZ sets the frequency to Hz range
FREQ_UNIT?
Returns the frequency unit currently being used by the frequency and pulse
modes.
FUNC?
Returns the current mode of the upper and lower displays. For example if
the calibrator is set to volts on the upper display, and pressure on the lower
display, FUNC? Would return:
DCV, PRESSURE
LOCAL
Restores the calibrator to local operation if it was in remote mode. Also
clears LOCKOUT if the unit was in lockout mode.
LOWER_MEAS
Sets the lower display to measure mode. The command is followed by any
of the parameters except for pulse and mA sim, which are source only. Enter
DCI for mA, DCV for volts, TC for thermocouple, RTD for RTD, FREQUENCY
for frequency, and PRESSURE for pressure. For example:
LOWER_MEAS DCV sets the lower display mode to VOLTS IN
L_PRES_UNIT
Sets the unit for measuring pressure on the lower display. Add the unit after
the command. The available pressure units and their syntax are shown in
the Table 7. (Parameter Units).
For example:
L_PRES_UNIT KPAL sets the pressure unit to kiloPascals
OUT
Sets the output of the calibrator. This command may be used to output mA,
volts, frequency, temperature, and ohms. Frequency output, which is set by
the command FREQ_TYPE, is either continuous or pulse. The calibrator is
43