TC_TYPE
Sets the type of the thermocouple. All available types are shown in the TC
Types table in Section 8. (Specifications). For example:
TC_TYPE B sets thermocouple type to B
TC_TYPE?
Returns the type of thermocouple the calibrator is set to.
TEMP_UNIT
Sets the temperature unit for sourcing and measuring RTD and TC. Add
CEL after the command for Celsius, and FAR for Fahrenheit. For example:
TEMP_UNIT CEL sets the temperature to be measured or sourced to
Celsius.
TEMP_UNIT?
Returns the temperature unit that is currently used for measuring and
sourcing RTD and TC.
TRIG
Starts and stops the pulse train when the calibrator is in pulse mode. The
parameters of the pulse train are set by commands PULSE_CNT, and
FREQ_LEVEL. Entering TRIG initializes the train. Entering the command
while the pulse train is running stops it.
TRIG?
Returns TRIGGERED if the pulse train is running, and returns
UNTRIGGERED when the pulse train is not running. Returns NONE when
the calibrator is not in pulse mode.
TSENS_TYPE
Sets the temperature sensor type to thermocouple, or to RTD for
temperature measurement. After the command add TC for thermocouple, or
RTD for RTDs. For example:
TSENS_TYPE TC sets the sensor type to thermocouple
46