Returns the status byte in decimal form from the Serial Poll Status Byte. For
example;
If 72 is returned, bits 6 and 3 are enabled.
*WAI
Prevents further remote commands from being executed until all previous
commands are executed. For example:
OUT 10 MA ; *WAI ; OUT 5 V will out put 10mA and wait until output
settles, than volts command will be processed.
7.5-2 Calibrator Commands
CAL_START
Puts the calibrator in calibration mode. The main display will say
CALIBRATION MODE and a calibration menu will be displayed on the
terminal.
CJC_STATE
Turns Cold Junction Compensation (CJC) on or off, when the calibrator is in
thermocouple (TC) mode. The command is used by adding ON or OFF after
it.
For example:
CJC_ STATE OFF turns CJC off.
CJC_STATE?
Tells whether the Cold Junction Compensation in thermocouple mode is
turned on or turned off. The calibrator returns OFF if CJC is off, and ON if
CJC is on.
CPRT_COEFA
This command is used for entering a custom RTD into the calibrator. The
numeric value entered after the command will be set as the first coefficient of
the polynomial used by the custom RTD.
For example:
CPRT_COEFA 3.9083E-3 enters 3.9083E-3 as coefficient A.