38
*ESR?
Returns the contents of the Event Status Register in decimal form. For
example:
If the ESR contains 10111001, *ESR? will return 185.
*IDN?
Returns the manufacturer, model number, and firmware revision of the
Calibrator. For example:
*IDN? will return AMETEK, ASC300, 250, 1.00
*OPC
Enables the Operation Complete setting in the ESR. This setting makes it
possible to check if an operations is complete after it has been initialized.
For example this operation could be used with the command TRIG.
*OPC?
Returns 1 when all operations are complete, and causes program execution
to pause until all the operations are complete. For example:
TRIG ; *OPC? will return a 1 when the pulse train initiated by TRIG is
complete.
*RST
Resets the state of calibrator to the power-up state. All subsequent
commands are held off until the execution of the command is complete.
*SRE
Loads a byte into the Service Request Enable register. A decimal number
must be entered, which when converted to binary, corresponds to the
correct settings.
For example:
*SRE 8 enters the binary number 00001000 to the SRE. This enables bit
3. Bit 6 is not used.
*SRE?
Returns a byte from the SRE. The byte is returned in decimal format. For
example:
If 40 is returned, bits 5 and 3 are enabled.