15
Caution
To avoid mechanically damaging the pressure module, never apply more than
13.5 Nm/10 ftlbs. of torque between the pressure module fittings, or between
the fittings an the body of the module.
To avoid damaging the pressure module from overpressure, never apply
pressure above the rated maximum printed on the module.
To avoid damaging the pressure module from corrosion, use it only with
specified materials. Refer to the pressure module documentation for material
compatibility.
To measure pressure, follow these steps:
1. Connect the pressure module to the calibrator as shown in Figure 9.
The calibrator can measure pressure on both the upper and the lower
display. This makes it possible to measure pressure in two different units
at the same time.
2. Switch to either upper or lower display from the Main Menu.
3. Select [PRESSURE] from the primary parameters.
4. Select the desired measuring unit.
5. Zero the pressure module. The zero function on the calibrator can be
found in the pressure zeroing menu.
Figure 9. Connections for Measuring Pressure
!