INPUT SPECIFICATIONS
ACCURACY IN C/F
±0.050°C(±0.090°F) @ 0°C(32°F)
±0.061°C(±0.110°F) @ 155°C(311°F)
±0.071°C(±0.127°F) @ 320°C(608°F)
±0.087°C(±0.156°F) @ 500°C(932°F)
Accuracy RTD Pt500
(excluding sensor accuracy)
±0.079°C(±0.142°F) @ -90°C(-130°F)
±0.083°C(±0.149°F) @ -50°C(-58°F)
±0.087°C(±0.157°F) @ 0°C(32°F)
±0.100°C(±0.180°F) @ 155°C(311°F)
±0.111°C(±0.200°F) @ 320°C(608°F)
±0.130°C(±0.235°F) @ 500°C(932°F)
Accuracy RTD Pt100
(excluding sensor accuracy)
±0.051°C(±0.092°F) @ -90°C(-130°F)
±0.054°C(±0.097°F) @ -50°C(-58°F)
±0.058°C(±0.104°F) @ 0°C(32°F)
±0.069°C(±0.124°F) @ 155°C(311°F)
±0.079°C(±0.142°F) @ 320°C(608°F)
±0.106°C(±0.191°F) @ 650°C(1202°F)
±0.112°C(±0.202°F) @ 700°C(1292°F)
Accuracy RTD Pt50
(excluding sensor accuracy)
±0.095°C(±0.171°F) @ -90°C(-130°F)
±0.098°C(±0.176°F) @ -50°C(-58°F)
±0.103°C(±0.185°F) @ 0°C(32°F)
±0.116°C(±0.209°F) @ 155°C(311°F)
±0.128°C(±0.230°F) @ 320°C(608°F)
±0.161°C(±0.290°F) @ 650°C(1202°F)
±0.169°C(±0.303°F) @ 700°C(1292°F)
Accuracy RTD Pt10
(excluding sensor accuracy)
±0.453°C(±0.815°F) @ -50°C(-58°F)
±0.462°C(±0.831°F) @ 0°C(32°F)
±0.495°C(±0.891°F) @ 155°C(311°F)
±0.524°C(±0.943°F) @ 320°C(608°F)
±0.610°C(±1.098°F) @ 650°C(1202°F)
±0.620°C(±1.116°F) @ 700°C(1292°F)