Caution…
 The well and the insertion tube must be clean before
use.
 Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes
carefully when not in use.
 The insertion tube must never be forced into the well.
The well could be damaged as a result, and the
insertion tube may get stuck.
Caution – Hot surface
 Do not touch the grid plate, the well or the insertion
tube while the calibrator is heating up – they may be
very hot.
 Do not touch the tip of the sensor when it is removed
from the insertion tube – it may be very hot.
 Do not touch the handle of the calibrator during use –
it may be very hot.
Caution – Cold surface
 If the calibrator has reached a temperature below
0°C/32°F, ice crystals may form on the insertion tube and
on the well. This, in turn, may cause verdigris to form on
the material.
To prevent this from happening, the insertion tube and the
well must be dried. This is done by heating up the
calibrator to min. 100°C/212°F and any water left will
evaporate.
Remove the insulation plug while heating up.
It is very important that humidity in the well and insertion
tube is removed to prevent corrosion and frost expansion
damages.
 Do not touch the well or insertion tube when these are
below 0°C/32°F – they can create frost-bites.

Table of Contents

Other manuals for Ametek JOFRA ATC-125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA ATC-125 and is the answer not in the manual?

Ametek JOFRA ATC-125 Specifications

General IconGeneral
BrandAmetek
ModelJOFRA ATC-125
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals