105446 06 23-03-2009 25
Silicone oil – Fill the well with oil according to the
tables of recommended oil volume. The recommended
volumes must be adjusted to the actual job.
For oil tables and further oil information – see section
3.3.2.
The sensor basket is marked with an optimum fluid level
mark (100%). When filling the well with fluid and placing
the sensors, this mark must never be exceeded.
Warning
Do not pour cold fluid into a hot well – it might cause an
explosion.
Do not pour water or any other fluids into a bath filled
with hot oil, because only a few drops of water might
cause a steam explosion, if poured into e.g. 250°C hot
oil.
Caution…
Do not use any alkali, acid or ionic fluids in the aluminium
well as it might be damaged.
Be careful not to overfill the well with oil.
The oil level rises several centimetres when the
temperature is rising to maximum. To stop the overflow
switch off the main power and the oil level will decent.
Remove excess hot fluid with the outmost care, as it might
be very hot.
Do not attempt to remove hot fluid with the liquid drainage
syringe, as it might melt.
Start the stirring magnet by following the procedure in section
4.8.7.
Place the calibration lid onto the well. See section 3.3.3 for
drilling information.
Select a SET-temperature according to the tables of
recommended oil volume by following the procedure in
section 4.4.