105446 06 23-03-2009 81
Nominal values Accepted range
 RTD and Reference Inputs:
50 (45 to 70)
176 (150 to 200)
345 (330 to 350)
 RTD inputs:
1000 (800 to 1200)
2700 (2300 to 2900)
 mA inputs:
0 mA (-1 to +1 mA)
10 mA (9 to 11 mA)
20 mA (18 to 22 mA)
 V Inputs:
0 V (-1 to +1 V)
5 V (3 to 7 V)
10 V (9 to 11 V)
 mV inputs:
0 mV (-1 to +1 mV)
35 mV (25 to 45 mV)
70 mV (63 to 77 mV)
 Cold Junction compensation: N-type thermocouple and a
reference junction, for example an
ice bath.
Note that Cold Junction compensation is only possible when
”Measure/Adjust/Calibrate” is selected.
The number and type of electrical inputs to be checked is user-
definable. You can also define if you want to check the inputs by just
measuring them or to measure, adjust and then calibrate the electrical
inputs.
Cold Junction compensation requires a special setup, using a
calibrated temperature source, for example an ice bath. This and the
measurement principles are shown briefly below: