AMETEK Denmark A/S does not take any
responsibility, if the well is filled with other fluids than
those recommended.
Liquid baths should only be operated by trained
personal.
Heat transfer fluids must only be used in calibrators
prepared as a liquid bath. If these fluids are overheated
they will create noxious or toxic fumes. Proper
ventilation must be used.
Product information on the fluid must be carefully
investigated before use.
Do not handle hot fluid.
If the oil is heated beyond the flash point, it may
constitute a fire hazard.
Do not pour water or any other fluids into a bath filled
with hot oil, because only a few drops of water might
cause a steam explosion, if poured into 250°C hot oil.
Do not under any circumstances pour water on burning
oil. It might cause a dangerous steam explosion.
Caution – Hot surface
This symbol is engraved in the grid plate.
Do not touch the grid plate, the well or the insertion
tube when the calibrator is heating up – they may be
very hot.
Do not touch the lid or the spill tray when the calibrator
is heating up – they may be very hot (ATC-140/250 A/B
only).
Do not touch the tip of the sensor when it is removed
from the insertion tube/well – it may be very hot.
Do not touch the handle of the calibrator during use – it
may be very hot.