123200 03 26-03-2004 11
• CTC-1200 A : ±0.8°C / ±1.44°F
There are two possible outcomes:
If the instrument detects excessive deviations for one or more steps,
the display will read =ERR. at the top. A message reading AdJ. will
flash at the bottom of the display to indicate that an adjustment is
required:
If, on the other hand, the deviations are found to be within the
permitted tolerances, the instrument will display the text =OK at the
top of the display. The text Cont. will flash at the bottom of the display
to indicate that you may continue without adjustments:
If the calibration results in the message ERR. AdJ., the following
options are available:
-
Press
to cancel the function and return to service mode.
-
Press
to go back and select between the calibration
temperatures if you want to repeat one of the steps.
-
Press
to toggle between ERR. AdJ. and ERR. Cont.