ECTROLAB — 23.07.2020 —
Malfunction Cause Remedy
Limit values are not
maintained during iCALising.
Insufficient intensity. 1. Check argon.
2. Clean spark stand,
adjust electrode.
3. Clean light inlet window.
4. Re-profile instrument
again and, if necessary,
contact the Service
Department.
During standardization,
factors < 2 are shown in red
or
Value dropped below the
minimum limit of the matrix-
depending reference
channel
e.g.: Fe 273 nm.
Insufficient intensity. 1. Check argon.
2. Clean spark stand,
adjust electrode.
3. Clean light inlet window.
4. Profile instrument,
sequence of
standardization
samples).
Call service if required.
During standardization,
factors < 0.5 are shown in
red
or
The maximum limit of the
typical measurement value
of a matrix-depending
reference channel has been
exceeded, e.g.: Fe 273 nm.
Intensities too high.
1. Check excitation
parameters.
Call service if required.