OM001-16 Otowave 102 Operating Manual Page 39
Guidance and manufacturer’s declaration electromagnetic immunity
(1)
The Otowave 102 Tympanometer is intended for use in the electromagnetic
environment specified below. The customer or user of the Otowave 102
Tympanometer should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
±1 kV
differential
mode
±2 kV
common
mode
Not applicable
Not applicable