Page 40 OM001-16 Otowave 102 Operating Manual
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0.5
cycle
40% U
T
(60% dip in
U
T
) for 5
cycles
70% U
T
(30% dip in
U
T
) for 25
cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
Not applicable
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to the application of the test level