OM001-16 Otowave 102 Operating Manual Page 41
Guidance and manufacturer’s declaration – electromagnetic immunity (2)
The Otowave 102 Tympanometer is intended for use in the electromagnetic
environment specified below. The customer or user of the Otowave 102
Tympanometer should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment guidance
Radiated
RF
IEC
61000-4-3
3 V/m
80MHz to
2.5GHz
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Otowave 102
Tympanometer, including
cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d = 1.2√P 80MHz to 800MHz
d = 2.3√P 800MHz to 2.5GHz
where P is the maximum
output power rating of the
transmitter in Watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,
a
should be less than
the compliance level in each
frequency range.
b