3-4 Commands Command Directory
3-4 PN: 10370-10374 Rev. F MG369xC GPIB PM
ADP1 <value> <unit>
Description: Opens the internal AM % depth parameter.
Changes the AM depth. Applies to internal linear AM.
Functional
Category: PARAMETER ENTRY
Value: 0% to 100%
Units: PCT
Example: AM7; ADP1 3 PCT
Turns on internal linear AM with a 3% depth.
Output Command: OAD1
Related
Commands: AM7, ADP2
Front Panel Key: Modulation | AM (linear) | Edit Depth
ADP2 <value> <unit>
Description: Opens the internal AM dB depth parameter.
Changes the AM depth. Applies to internal log AM.
Functional
Category: PARAMETER ENTRY
Value: 0 dB to 25 dB
Units: DB
Example: AM8; ADP2 3 DB
Turns on internal log AM with a 3 dB depth.
Output Command: OAD2
Related
Commands: AM8, ADP1
Front Panel Key: Modulation | AM (log) | Edit Depth
ADR
Description: GPIB address terminator.
Functional
Category: DATA TERMINATOR
Front Panel Key: N/A