Setting Up Markers
Spectrum Master UG PN: 10580-00277 Rev. E 3-9
3-7 Setting Up Markers
Press the Marker main menu key to display the Marker menu. Refer to Figure 3-3.
Selecting, Activating, and Placing a Marker
1. Press the Marker123456 submenu key so that the desired marker number is
underlined. Each press of the Marker123456 submenu key toggles to the next higher
number (cycling from 6 back to 1 again). Press Shift, then the Marker123456
submenu key to cycle backward through the marker list.
2. Press the On Off submenu key so that On is underlined. The selected marker is
displayed in red, ready to be moved.
3. Use the rotary knob to place the marker on the desired frequency.
4. Repeat Step 1 and Step 2 to activate and move multiple markers.
Selecting, Activating, and Placing a Delta Marker:
1. Press the Marker123456 submenu key so that the desired delta marker number is
underlined.
2. Press the Delta On Off submenu key so that On is underlined. The selected marker is
displayed in red, ready to be moved.
3. Use the rotary knob or the numeric keypad to place the delta marker on the desired
offset frequency from the associated reference marker.
4. Repeat Step 1 and Step 2 to activate and move multiple markers.
Viewing Marker Data in a Table Format
1. Press the More submenu key.