Spectrum Master UG PN: 10580-00277 Rev. E B-1
Appendix B — Error Messages
B-1 Introduction
This chapter provides a list of information and error messages that could be displayed on the
MS272xC Spectrum Master. If any error condition persists, then contact your local Anritsu
Service Center (http://www.anritsu.com/Contact.asp).
B-2 Self Test or Application Self Test Error Messages
Overall Status FAILED
One or more elements of the System or Application Self Test has failed. Refer to the other
pass fail tests listed below to determine which specific test failed.
ADC Self Test FAILED
The Analog to Digital converter failed to return an answer. Insure that the battery level is
adequate for operation or that the temperature is within acceptable limits. Reset to factory
defaults with either Factory Defaults, ESC+ON, or MASTER RESET, System+ON.
Caution: Use of MASTER RESET, System+ON, will erase all user saved setups and
measurement traces and return the unit to a fully Factory Default condition. If the error
persists, then contact your Anritsu Service Center.
DDC FAILED
The Digital Down Converter failed to return a value. Insure that the battery level is adequate
for operation or that temperature is within acceptable limits. Reset to factory defaults with
either Factory Defaults, ESC+ON, or MASTER RESET, System+ON. Caution: Use of
MASTER RESET, System+ON, will erase all user saved setups and measurement traces and
return the unit to a fully Factory Default condition. If the error persists, then contact your
Anritsu Service Center.
Lock Test FAILED
One or more Phase Lock Loops Failed to properly achieve Lock Status. Insure that the
battery level is adequate for operation or that temperature is within acceptable limits. Reset
to factory defaults with either Factory Defaults, ESC+ON, or MASTER RESET, System+ON.
Caution: Use of MASTER RESET, System+ON, will erase all user saved setups and
measurement traces and return the unit to a fully Factory Default condition. If the error
persists, then contact your Anritsu Service Center.
Over Power FAILED