Disinfectant fluid tank filling
To fill up, use pure PEROXY Ag+ or 3% oxygenated water (10 volumes) only, without diluting them.
Dental units of A5 and A6 PLUS model
• Open the water unit side cover.
• Turn the tank (a).
• Remove plug (k) and pour disinfectant fluid inside tank until it is full.
The plug shape allows it to be used as a funnel to make filling
easier.
• Refit plug and tank.
• Lastly, close the water unit side cover.
Dental units of L6 model
• Open tank flap.
• Remove plug (k) and pour disinfectant fluid inside tank until it is full.
The plug shape allows it to be used as a funnel to make filling
easier.
• Refit plug and close tank flap.
Emptying the W.H.E. system water circuit
This function allows emptying the W.H.E. system water circuit in case the
dental unit must stay off for several days.
Proceed as follows:
• Place the supplied container (e) under cup spout and enable the water
circuit emptying cycle by keeping the button (H) pressed for at least 5
seconds; the display shows the system’s code.
• Wait for the water circuit to be emptied, a sound (3 BEEPS) will be
emitted at the end of the operation.
• Now the dental unit is in lockout mode and it can be turned off.
At restart, theW.H.E. system will be automatically restored.
Warning messages on Display.
Console with Digit DISPLAY:
If the system detects a fault operation, it emits an intermittent sound (2 BEEPS) and locks out, preventing water from being delivered to the dental unit.
In these cases, it is recommended to first check the correct supply of water and air to the dental unit.
Then turn the dental unit off and on again, if the lockout condition persists, call the Technical Service department.
Console with LCD Touch DISPLAY:
If system detects a malfunction, an error message will be displayed (see paragraph 10).
If the detected error is negligible, the dental unit continues working, while in case of severe errors, the dental unit will be blocked and Technical Service is
required.
PEROXY Ag+ storage.
For proper PEROXY Ag+ storage, follow the manufacturer’s instructions given on the package.
It is important to keep the package tightly closed and stored in a cool place at a temperature not exceeding 25°C.
Never leave PEROXY Ag+ or oxygenated water in the tank (a) for more than one month.
If you are going to be absent from the surgery for long periods of time (holidays), completely empty the tank (a) before leaving.
To empty tank, use a suction tube.