10. WARNING/ERROR MESSAGES
Two types of notices can be displayed on the console display: warning messages (Wxxx) and error messages (Exxx).
Warning messages (Wxxx) indicate situations that require user intervention.
Error messages (Exxx) indicate situations that cannot be fixed by the user and require the intervention of the Technical Service.
When contacting the Technical Service, please communicate the error number displayed.
Warning
Cause
Solution
W050
The disinfection cycle preparation page opens with the
instruments not in rest position.
To start a disinfection cycle, all disinfection instruments must be
in a rest position.
W051
BIOSTER S start request performed without selecting any
instrument or cup.
Select the instrument, cup or suction tubes, then activate the
disinfection cycle.
W057
The system has stopped the disinfection cycle after detecting
instruments changed during the cycle.
To restart the disinfection cycle after a lockout, use the same
instruments extracted during the stopped cycle.
W059
BIOSTER S start request performed without any extracted
instrument or selected cup.
Select the instrument or cup, then start the disinfection cycle.
W060
H2O2 tank empty.
Fill H2O2 tank.
W061
A user has requested to skip the disinfection pause.
None
W062
At the end of the cycle, a warning is issued if the pause has not
been made.
None
W063
H2O tank empty.
Fill H2O tank.
W064
H2O2 tank empty.
Fill H2O2 tank.
W065
Dentist’s board water cocks closed.
Check that the dentist’s board water cocks are open, if the
problem persists, contact technical service department.
W066
Emptying is not possible since bypass is not selected.
Select mains water supply.
W0100
An instrument has been extracted with active foot control.
Release the foot control, then activate the instrument.
W0101
Instruments extracted at start-up.
Put the instruments back in place.
W102
Instrument not detected.
Check connections, if the problem persists, contact the technical
service department.
W103
A Flushing request has been made with no extracted instrument.
Repeat the Flushing cycle request with instrument extracted.
W104
A Flushing request has been made with SHS tank in reserve.
Fill H2O tank and repeat the Flushing request, if the problem
persists, contact technical service department.
W105
Flushing Ended with error.
Repeat the Flushing cycle.
W106
Flushing is not possible since bypass with long cycle is not
selected.
Repeat the Flushing cycle request with mains water selected.
W150
Suction tubes extracted at start-up.
Put suction tubes back in place.
W151
Instruments extracted at start-up.
Put instruments back in place.