7.4. BIOSTER S AUTOMATIC DISINFECTION SYSTEM (only dental units with LCD Touch DISPLAY)
Description of the system.
This system allows carrying out an automatic disinfection cycle of the
water circuits of the following instruments:
• All instruments that require water for their use are placed on the dentist's
board.
• The syringe and any other instruments that require water for their use
are placed on the assistant's board.
• Water-to-cup ducts.
In addition, system features a tank (a) positioned inside the water unit and
suitable to contain approx. 500 cc. of disinfectant fluid.
Carry out a disinfection cycle at the end of every working
day.
Disinfectant fluid running out warning.
When the disinfectant fluid present inside tank (a) is running out, a special
warning icon (A) appears on the DISPLAY, an error message appears on
the DISPLAY and 3 warning BEEPS are emitted and repeated whenever
dental unit is turned on.
Disinfectant fluid tank filling.
See paragraph 7.3.
To fill up, use pure PEROXY Ag+ or 3% oxygenated water
(10 volumes) only, without diluting them.
Setting the disinfection cycle.
• Check disinfectant fluid level inside tank and, if necessary, top up.
The disinfection cycle will not be activated if the fluid level
inside tank is below reserve level.
• Use the button "BIO" on the assistant's board to access the "BIOSTER
S disinfection cycle setting" menu and set the residence time of the
disinfectant fluid inside instrument water ducts (see paragraph 5.1.1.1).
• Position the special container (d) for the instruments to be disinfected
on the bowl.
• Insert the hoses of the instruments to be disinfected in the container
For the syringe instrument, it is necessary to remove the
handle and completely insert the dedicated adapter (f)
(second click). The heating system must be switched off.
The micromotor hose must be fully inserted inside motor
body.
Turbine and scaler hoses must be inserted without the
handpiece.
• Should you wish to disinfect suction tubes, insert tube terminals inside
the special couplings under manifold (see paragraph 7.5.).
Make sure that suction tube terminals are open.
• If water-to-cup duct disinfection is selected, place the special supplied
cup (e) under cup spout.
• Make sure that spray cocks (g), positioned on board lower side, are fully
open.