A5 - A6 PLUS - L6
98
EN
WHEN
PART
ACTION
REFERENCE
PARAGRAPH
When needed.
Removable instrument hoses.
Clean with suitable disinfectant in compliance with
manufacturer's recommendations.
Spray product on disposable soft paper.
Do not use acids or abrasive products.
See paragraph 5.
Spray ducts.
Disinfection of spray water ducts with manual cycle (if
the S.H.S. system is fitted).
See paragraph 7.2.1.
Disinfection of spray water ducts with semi-automatic
cycle (if the system is fitted).
See paragraph 7.2.
Air-liquid separator canister by
CATTANI
Clean the separator’s container, drain valve and
probes.
See paragraph 9.5.
METASYS amalgam separator.
Empty separator container.
See documentation
enclosed with separator.
DÜRR amalgam separator.
Empty separator container.
See documentation
enclosed with separator.
Operating light
Clean the front lens and the reflector
See documents enclosed
with equipment.
VENUS PLUS-L LED operating light
Clean transparent screen and handles
See paragraph 8.1.
Monitor on light pole
Clean surfaces
See documents enclosed
with equipment.
Coated surfaces and upholstery.
Clean with suitable disinfectant in compliance with
manufacturer's recommendations.
Spray product on disposable soft paper.
Do not use acids or abrasive products.
See paragraph 1.4.
Weekly.
Suction tube holder terminals.
Lubricate the O-ring.
See paragraph 9.4.
Monthly.
Turbine return air filter.
Check the filter and replace it if necessary (code
97290014).
See paragraph 9.6.
Yearly.
Dental chair and dental unit.
Contact the Technical Service for general inspection.
/