EasyManua.ls Logo

Apera Instruments 800 Series - P1.2 - Set up calibration reminder; P1.4 - Set the Factory Default (No-Yes); 8.4 pH Setting Sub-Menu

Apera Instruments 800 Series
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 24 -
P1.2 — Set up calibration reminder (No—H00—D00)
1. Press , ”No” flashes, press to choose No→H00→D00;
NO—no setting; H00—set hours(0~99); D00—set days(0~99).
2. When “H” flashes, press , “00” flashes, press to adjust hours,
press to confirm; When “D” flashes, press ,“00”flashes, press
to adjust days, press to confirm; When “No” flashes, press
to confirm.
3. Press to enter P1.3, or press to return to measurement mode.
P1.3 — Check Calibration Time
1. The data and time of last calibration displayed on the left image is: On
June 8, 2016 to 9:08
2. Press to enter P1.4, or press to return to measurement mode.
P1.4 — Set the Factory Default(No—Yes)
Press , “NO” flashes, press to select NO→Yes, press to
confirm, instrument back to measurement mode.
No — Not restore; Yes — Restore factory default
8.4 pH Setting Sub-Menu (press or to switch)
P2.1 — Select electrode’s constant(1.0—10.0—0.1)
1. In P2.0, press to enter P2.1
2. Press , “1.0” flashes, press to choose 1.0→10.0→0.1; press
to confirm.
3. Press to enter P2.2, or press to return to measurement mode.
P2.2 — Set up calibration reminder (No—H00—D00)
1. Press , ”No” flashes, press to choose No→H00→D00;
NO—no setting; H00—set hours(0~99); D00—set days(0~99).
2. When “H” flashes, press , “00” flashes, press to adjust hours,
press to confirm; When “D” flashes, press ,“00”flashes, press
to adjust days, press to confirm; When “No” flashes, press
to confirm.
3. Press to enter P2.3, or press to return to measurement mode.
P2.3 — Check Calibration Time
1. The data and time of last calibration displayed on the left image is: On
August 6, 2016 to 12:00.
2. Press to enter P2.4, or press to return to measurement mode.