SPECIFICATIONS
Thermolysis:
PicoFlash, MeloFlash, MultiPlex, Synchro*
(*included in the xCell Pro version, optional for the
xCell Pure version)
ISM Frequency: 27.12 MHz
Heating intensity (MultiPlex):
adjustable from 0 to 29%, in 1% increments
Pulse intensity (PicoFlash, MeloFlash, MultiPlex,
Synchro):
adjustable from 0 to 100%, in 1% increments
Pulse duration (PicoFlash, MultiPlex):
adjustable from 0.000 to 0.500s,
in 0.001s increments
Pulse duration (MeloFlash, Synchro):
adjustable from 0.00 to 0.99s, in 0.01s increments
Heating duration (MultiPlex):
adjustable from 0.0 to 10.0s
Number of impulses (PicoFlash, MultiPlex):
adjustable from 0 to 9 in increments of 1
Proportional Thermolysis:
active in PicoFlash, MeloFlash and MultiPlex
Indicator light: when thermolysis current is active
Audible notification: at the end of an impulse
Anaphoresis or cataphoresis
Regulated current:
adjustable from 0.00 to 2.00mA (anaphoresis) or
0.00 or 3.00mA (cataphoresis), in 0.01mA incre-
ments
Polarity: electronic inversion
Indicator light: current flow
Cataphoresis programs: 110
Combined current (Blend):
PicoBlend, OmniBlend, MultiBlend, SynchroBlend*,
EvoluBlend* (*included in the Pro version, optional
for the Pure version)
Galvanic current:
adjustable from 0.00 to 0.99mA, in 0.01mA incre-
ments
Treatment duration: adjutable from 0 to 30s
(OmniBlend, MultiBlend, EvoluBlend), from 0 to
60s (PicoBlend),(SyncroBlend) from 4 to 60s in 1s
increments
Thermolysis impulse intensity (PicoBlend):
adjustable from 0 to 29%, in 1% increments
Number of impulses (PicoBlend):
adjustable from 0 to 9 in increments of 1
Maximum intensity of thermolytic heating
(MultiBlend, EvoluBlend, OmniBlend):
adjustable from 0 to 29%, in 1% increments
Final thermolysis impulse intensity (MultiBlend):
adjustable from 0 to 100%, in 1% increments
Intensité de l’impulsion (SynchroBlend) :
adjustable from 0 to 10%, in 1 % increments
Duration of the final impulse (MultiBlend):
adjustable from 0.000 to 0.500s in 0.001s incre-
ments
U.L. (Units of Lye): from 0 to 594
The application of currents:
switchable - can begin with galvanic or can begin
with thermolysis
Indicator light:
galvanic current is active and thermolysis current
is active
Audible notification:
at the end of an impulse