51
© 2014 Dectro International. All rights reserved.
GUIDE AND MANUFACTURER'S DECLARATION -
ELECTROMAGNETIC IMMUNITY
The Apilus xCell is for use in the electromagnetic environment as described below.
The customer or the user of the Apilus xCell must ensure that it is used in such an environment.
Immunity Test
Test Condtion
CEI 60601
Level of Compliance Electromagnetic Environment Guide
Electrostatic
Discharge (ESD)
CEI 61000-4-2
± 6kV Contact
± 8kV Air
± 6kV Contact
± 8kV Air
Floors should be covered with wood,
concrete or ceramic. When the floors
are covered with synthetic material,
the relative humidity should be main-
tained at more than 30%.
Transient
fast/surge
CEI 61000-4-4
± 2kV
Supply lines
± 2kV
Supply lines
The quality of the voltage distribution
must be that of the environment of a
typical commercial or hospital.
Surge
CEI 61000-4-5
± 1kV Between lines
± 2kV Between each line
and grounding
± 1kV Between lines
± 2kV Between each line
and grounding
The quality of the voltage distribution
must be that of the environment of a
typical commercial or hospital
Voltage dips, short
interruptions and
voltage variations
CEI 61000-4-11
<5 % U
T
(>95 % reduction de U
T
)
pour 0.5 cycle
40 % U
T
(60 % reduction de U
T
)
for 5 cycle
70 % U
T
(30 % reduction de U
T
)
for 25 cycle
<5 % U
T
(>95 % reduction de U
T
)
for 5s
<5 % U
T
(>95 % reduction de U
T
)
pour 0.5 cycle
40 % U
T
(60 % reduction de U
T
)
for 5 cycle
70 % U
T
(30 % reduction de U
T
)
for 25 cycle
<5 % U
T
(>95 % reduction de U
T
)
for 5s
The quality of the voltage distribution
must be that of the environment of
a typical commercial or hospital. It
is recommended that you connect
Apilus xCell to an uninterruptible
power supply or a battery if the user
requires continuous operation during
power interruptions distribution.
Magnetic fields
(50/60Hz)
CEI 61000-4-8
3A/m 3A/m
Magnetic fields should be at a typical
level as found in a typical environ-
ment of a business or a hospital.
NOTE : U
T
is the distribution voltage prior to application of the test level.