A1 24 DTR.APC.APR.ALW.03
(ENG)
5.7. APC … APR … Environmental parameters
5.7.1. Electromagnetic Compatibility (EMC),
Immunity
Criterion according to EN 61326-1,2
Electrostatic Discharge (ESD):
EN 61000-4-2
level 3,
contact ±6kV
air ±8kV
Criterion A
Conducted Radio Frequency:
EN 61000-4-6
0,15… 80MHz, 10V
Criterion A
Radiated Electromagnetic Field:
EN 61000-4-3
80… 2 000MHz – 10V/m
… 2 700MHz – 1V/m
Criterion A
Electrical Fast Transient (Burst):
EN 61000-4-4
± 1kV
Criterion A
Electrical Slow Transient (Surge):
EN 61000-4-5
± 1kV
Criterion B
5.7.2. Electromagnetic Compatibility,
emission
Criterion according to CISPR16-1, CISPR 16-2, class B
Radiated Emission:
Distance from antenna: 3m
limits quasi-peak:
0,15 … 30MHz, 80-52dBV/m;
30 … 2000MHz, <54dBV/m
Conducted Emission:
limits quasi-peak:
0,01 … 0,150MHz, 96-50dBV/m;
0,150 … 0,350MHz, 60-50dBV/m;
0,35 … 30MHz, <50dBV/m
5.7.3. Climatic Immunity
Temperature:
EN 60068-2-1, EN 60068-2-2
hot: T = 55
0
C, RH = max 55%
cold: T = -25
0
C,
Damp Heat Cycle:
EN 60068-2-30,
(T = 55
0
C, RH = min95%, 24h)x2
Salt Mist:
5% NaCl, pH 6,5 … 7.2 at 20
0
C
T = 40
0
C, RH = min93%, 28dni
5.7.4. Mechanical Immunity
Shocks:
EN 60068-2-27
50g/11ms
Sinusoidal Vibrations:
EN 60068-2-6, test Fc
do 1,6mm, 2 … 25Hz
do 4g for 25 …100Hz
5.7.5. Electrical Isolation
>100 M @750V DC Ex
>100 M @110V DC normal, marine, PED
5.7.6. Insulation Strength
550V AC or 750V DC, 1min
Ex, marine
75VAC or 110V DC, 1min
normal, PED
5.7.7. Enclosure Ingress Protection
EN 60529
IP 66,67