Safety and EMC Compliance Information
xiv GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Electrical Safety
Fuses
ELECTRICAL SHOCK HAZARD. Improper fuses or high-
voltage supply can damage the instrument wiring system and cause a fire. Before
turning on the instrument, verify that the fuses are properly installed and that the
instrument voltage matches the power supply in your laboratory.
FIRE HAZARD. For continued protection against the risk of
fire, replace fuses only with fuses of the type and rating specified for the instrument.
Power
ELECTRICAL HAZARD. Grounding circuit continuity is vital
for the safe operation of equipment. Never operate equipment with the grounding
conductor disconnected.
ELECTRICAL HAZARD. Use properly configured and
approved line cords for the voltage supply in your facility.
ELECTRICAL HAZARD. Plug the system into a properly
grounded receptacle with adequate current capacity.
Overvoltage
Rating
The GeneAmp PCR System
®
9700 system has an installation (overvoltage) category
of II, and is classified as portable equipment
Physical Hazard Safety
Moving Parts
PHYSICAL INJURY HAZARD. Moving parts can crush and
cut. Keep hands clear of moving parts while operating the instrument. Disconnect
power before servicing the instrument.
PHYSICAL INJURY HAZARD. Do not operate the instrument
without the arm shield in place. Keep hands out of the deck area when the instrument
is spotting.
Hot Surface
PHYSICAL INJURY HAZARD. During instrument operation,
the temperature of the heated cover can be as high as 108 °C, and the temperature of
the sample block can e as high as 100 °C. Before performing the procedure, keep
hands away until the heated cover and sample block reach room temperature.

Other manuals for Applied Biosystems GeneAmp PCR System 9700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems GeneAmp PCR System 9700 and is the answer not in the manual?

Applied Biosystems GeneAmp PCR System 9700 Specifications

General IconGeneral
BrandApplied Biosystems
ModelGeneAmp PCR System 9700
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals