Chapter 1 Dual 96 Sample Block Module Overview
1-2 GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
About the Dual 96 Sample Block Module
What It Does
The GeneAmp
®
PCR System 9700 amplifies nucleic acids using the polymerase
chain reaction (PCR) process.
Sample block modules are interchangeable. You can, for example, remove a 384-well
Sample Block Module and replace it with the Dual 96 Sample Block Module. This
interchangeable module design allows you to change sample well formats as well as
throughput capacity.
Compatible
Software
Interchanging sample block modules may require a firmware change. If the sample
block module is placed on a base module running an earlier software version, the
instrument will not operate correctly. To upgrade the base module software, use the
floppy disk that is shipped with the Dual 96 Sample Block Module. For instructions
on upgrading the software, see the GeneAmp PCR System 9700 Base Module User’s
Manual (P/N 4303481) or see “How to Obtain Services and Support”on page vi.
Heated Cover
The Dual 96 Sample Block Module comes with a heated manual lid. Simply slide the
lid back and forth into position. The heated cover slides over the sample blocks and
performs two functions:
• It raises the temperature of the upper part of the tube (the part above the sample
blocks) above the temperature of the reaction mixture. This reduces
condensation, which may have a negative effect on your chemistry.
• It puts pressure on the reaction plates to seat them firmly and precisely in the
sample blocks. This enhances heat transfer.