Chapter 3 Testing the Dual 96 Sample Block Module
3-14 GeneAmp PCR System 9700 Dual 96 Sample Block Module User Guide
Data Sheet:
Temperature
Non-Uniformity
Te st
When you run the Temperature Non-Uniformity Test, record the setpoint values for
wells L1–L6 and R1–R6 on a copy of this data sheet. At the end of the Temperature
Non-Uniformity Test, check the values displayed on the System 9700 against the
values recorded to ensure accurate test records.
IMPORTANT! Use a photocopy of this page to record values, and keep the original
intact.
Date
Te s t e d B y
Probe Serial No.
Meter Serial No.
Setpoint Value 95°C 55°C
L1
L2
L3
L4
L5
L6
R1
R2
R3
R4
R5
R6