Note: Netw
orked (LAN) congurations can be internal congurations that do not
require internet access. However, if you are networking an instrument for Cloud
access, the LAN must have internet access.
IMPORTANT! For all wired congurations using an Ethernet cable, we support
DHCP or static IP only
. Proxy servers are not supported.
A networked computer can detect all instruments that are on the same netw
ork
and subnet mask.
However, the computer can only control one instrument at a time.
All networked computers can detect each instrument that is on the same network
and subnet mask.
However, the instrument can only be controlled by one computer at a time.
Note: For detailed information about networking your instrument, see the
QuantStudio
3 and 5 Real-Time PCR Systems IT Checklist (Pub. No. MAN0013738).
The instrument:
Is factory-congured for IPv4 TCP/IP communication and includes an Ethernet
adapter (100/1,000 Mbps) with an RJ45-type connector for integrating the device
into a local area network (LAN).
Can alternatively be congured for wireless networking (High Power USB WiFi
Module required, sold separately as an optional accessory).
The instrument can be congured for either wired or wireless networking, not both.
If a Thermo Fisher Scientic service representative is to install the instrument:
If the instrument will be connected to a LAN, an active, tested network jack must
be in place before the scheduled installation date.
A representative from your information technologies department must be
available during the installation to help connect the instrument to your network.
Required materials to network the instrument:
Wired—Ethernet cable of sucient length with RJ45 connectors
CAT5 cable for a 100 Mbps network connection
CAT5e or CAT6 cable for a 1,000 Mbps network connection
or
Wireless—High Power USB WiFi Module (Cat. No. A26774, sold separately)
Ports
Condition
80/443
Standard ports for
ins
trument-to-Cloud and computer-to-Cloud connections
mDNS, 7000 Instrument-to-computer connection
mDNS, 5353 Instrument discovery
Networked
(local area
network– LAN)
configuration
Network
r
equirements
Firewall ports that
must be open
Instrument-to-computer connections
QuantStudio
3 and 5 Real-Time PCR Systems Site Preparation Guide | MAN0010405 C.0
11

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems QuantStudio 5 and is the answer not in the manual?

Applied Biosystems QuantStudio 5 Specifications

General IconGeneral
BrandApplied Biosystems
ModelQuantStudio 5
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals