Fixed Channel - channel selection for fixed mode – one of 6 options (combinations L1 – L3 and
voltage, current). However, the previous parameter „Channel Mode” has to be
set to FIXED.
Remark: If the value is placed in parentheses, it is not considered, because the superior parameter is
most likely set into AUTO option.
3.2.10 History Mode Menu
Show History -> - display of historical statistics
State - enabling or disabling the History function
Quantities - without the possibility of any change (all quantities)
Integr. Time - integration time is always equal to the value of TIME BASE, which is set in
Measurement Menu. INTEGR. TIME item does not allow to change its value.
Reset - resetting the History records
3.2.11 Menu TALK ONLY
State - enabling or disabling TALK ONLY Mode, which serves to continuous sending
of measured data to communication port
COM1 ... - to enter the settings of COM1 port in order to get more options of TALK ONLY
Mode
3.2.12 Menu TALK ONLY COM1
State - enabling or disabling TALK ONLY Mode on COM1 port
Format - sent data format setting either to SHORT or LONG
Quantities ... - to enter the list of quantities where every single available quantity can be
selected in order to continuously send their values via COM1 port
3.2.13 Output Frequency Response Menu
Quick Mode - enabling or disabling the Quick Response Mode
Threshold - setting the threshold value in %