5.17.3 Commands
SYSTem:TALKonly {0 | 1}
General disabling (0) or enabling (1) of ‘TALK ONLY’ mode.
SYSTem:TALKonly?
Request for state of ‘TALK ONLY’ mode. Reply is 0 if disabled or 1 if enabled.
SYSTem:TALKonly:COM1 {0 | 1}
Disabling (0) or enabling (1) of COM1 (RS-232) output in ‘TALK ONLY’ mode.
SYSTem:TALKonly:COM1?
Request for state of COM1 (RS-232) output in ‘TALK ONLY’ mode. Reply is 0 if disabled or 1 if enabled.
SYSTem:TALKonly:COM1:FORMat {0 | 1}
Setting LONG format (0) or SHORT format (1) of COM1 output in ‘TALK ONLY’ mode.
SYSTem:TALKonly:COM1:FORMat?
Request for format setting of COM1 output in ‘TALK ONLY’ mode. Reply is 0 for LONG format or 1 for
SHORT format.
SYSTem:TALKonly:COM1:QUANtities {<qParam>}
Setting of quantities sent on COM1 output in ‘TALK ONLY’ mode. Quantities parameter (qParam) is
decimal number containing binary information about enabled/disabled quantities. For format of
parameter see detailed description in Mode ‘TALK ONLY’ chapter.
SYSTem:TALKonly:COM1:QUANtities?
Request for setting of quantities sent on COM1 output in ‘TALK ONLY’ mode. Reply is decimal number
containing binary information about enabled/disabled quantities. For format of parameter see detailed
description in Mode ‘TALK ONLY’ chapter.
5.17.4 Quantities parameter
Quantities parameter is 32-bit decimal word, used in commands for enabling/disabling of quantities.
Every quantity Qx is represented by bit X. Value 0 of this bit X disables quantity Qx and value 1 of bit
X enables quantity Qx.
Format of quantities parameter:
bit #
21
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
Q #
Q21
Q20
Q19
Q18
Q17
Q16
Q15
Q14
Q13
Q12
Q11
Q10
Q09
Q08
Q07
Q06
Q05
Q04
Q03
Q02
Q01
Example: For enabling Q01(Active Power), Q02(Voltage), Q03(Current) and Q11(Frequency) has
quantity parameter value 2055 (= 2
1
+ 2
2
+ 2
3
+ 2
11
).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Precision RS 2130 and is the answer not in the manual?

Applied Precision RS 2130 Specifications

General IconGeneral
BrandApplied Precision
ModelRS 2130
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals