Reference Standard RS 2x30, RS 1x30
User’s Guide
RS-UG91-EN www.appliedp.com
48
5.18.6 Controlling and configuring of dosage function
Dosage function can be controlled and configured by submenu “Measurement / Energy / Dosage” or
by remote commands:
Commands for controlling of dosage function:
SYSTem:ENERgy:DOSage:INIT
Initialization of dosage function.
SYSTem:ENERgy:DOSage:STARt
Start (or initialization) of dosage function.
SYSTem:ENERgy:DOSage:STOP
Stop (or cancel) of dosage function.
SYSTem:ENERgy:DOSage:CANCel
Cancel of dosage function.
Commands for configuring dosage:
SYSTem:ENERgy:DOSage:MODe {<mode>}
Setting of Dosage Mode. Parameter is number (0-5) or code (TIME, ACTIVE …) of Dosage Mode.
SYSTem:ENERgy:DOSage:MODe?
Request for state of Dosage Mode.
SYSTem:ENERgy:DOSage:TIMe {<time>}
Setting of time (in seconds) for time dosage.
SYSTem:ENERgy:DOSage:TIMe?
Request for time for time dosage.
SYSTem:ENERgy:DOSage:ENERgy {<energy>}
Setting of energy (in Ws or VAs) for energy dosage.
SYSTem:ENERgy:DOSage:ENERgy?
Request for energy (in Ws or VAs) for energy dosage.
SYSTem:ENERgy:DOSage:ENERgy:KWH {<energy>}
Setting of energy (in kWh or kVAh) for energy dosage.
SYSTem:ENERgy:DOSage:ENERgy:KWH?
Request for energy (in kWh or kVAh) for energy dosage.
Commands for configuring SU/SI contacts mode:
SYSTem:ENERgy:DOSage:CONTact:VOLTage:MODe {<mode>}
Setting of SU contact mode. Parameter <mode> is 0(contact always OPEN), 1(contact always
CLOSED) or 2(contact AUTOmatically controlled by device).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:MODe?
Request for setting of SU contact mode. Reply is 0(contact always OPEN), 1(contact always
CLOSED) or 2(contact AUTOmatically controlled by device depending on configuration).
SYSTem:ENERgy:DOSage:CONTact:CURRent:MODe {<mode>}
Setting of SI contact mode. Parameter <mode> is 0(contact always OPEN), 1(contact always
CLOSED) or 2(contact AUTOmatically controlled by device).
SYSTem:ENERgy:DOSage:CONTact:CURRent:MODe?
Request for setting of SI contact mode. Reply is 0(contact always OPEN), 1(contact always CLOSED)
or 2(contact AUTOmatically controlled by device depending on configuration).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Precision RS 2130 and is the answer not in the manual?

Applied Precision RS 2130 Specifications

General IconGeneral
BrandApplied Precision
ModelRS 2130
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals