Reference Standard RS 2x30, RS 1x30
User’s Guide
www.appliedp.com RS-UG91-EN
49
Commands for configuring SU/SI contacts behavior in AUTO mode:
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STARt:STATe {<state>}
Setting of SU contact state in AUTO mode after start of dosage function. Parameter <state> is
0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STARt:STATe?
Request for setting of SU contact state in AUTO mode after start of dosage function. Reply is
0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STARt:STATe {<state>}
Setting of SI contact state in AUTO mode after start of dosage function. Parameter <state> is
0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STARt:STATe?
Request for setting of SI contact state in AUTO mode after start of dosage function. Reply is 0(contact
OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STOP:STATe {<state>}
Setting of SU contact state in AUTO mode before start and after end of dosage function. Parameter
<state> is 0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STOP:STATe?
Request for setting of SU contact state in AUTO mode before start and after end of dosage function.
Reply is 0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STOP:STATe {<state>}
Setting of SI contact state in AUTO mode before start and after end of dosage function. Parameter
<state> is 0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STOP:STATe?
Request for setting of SI contact state in AUTO mode before start and after end of dosage function.
Reply is 0(contact OPEN), 1(contact CLOSED) or 2(contact state UNCHANGED).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STARt:DELay {<delay>}
Setting of SU contact switch delay in AUTO mode after start of dosage function. Parameter <delay> is
delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STARt:DELay?
Request for setting of SU contact switch delay in AUTO mode after start of dosage function. Reply is
delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STARt:DELay {<delay>}
Setting of SI contact switch delay in AUTO mode after start of dosage function. Parameter <delay> is
delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STARt:DELay?
Request for setting of SI contact switch delay in AUTO mode after start of dosage function. Reply is
delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STOP:DELay {<delay>}
Setting of SU contact switch delay in AUTO mode before start and after end of dosage function.
Parameter <delay> is delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:VOLTage:AUTo:STOP:DELay?
Request for setting of SU contact switch delay in AUTO mode before start and after end of dosage
function. Reply is delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STOP:DELay {<delay>}
Setting of SI contact switch delay in AUTO mode before start and after end of dosage function.
Parameter <delay> is delay in seconds (e.g. 0.05).
SYSTem:ENERgy:DOSage:CONTact:CURRent:AUTo:STOP:DELay?
Request for setting of SI contact switch delay in AUTO mode before start and after end of dosage
function. Reply is delay in seconds (e.g. 0.05).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Precision RS 2130 and is the answer not in the manual?

Applied Precision RS 2130 Specifications

General IconGeneral
BrandApplied Precision
ModelRS 2130
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals