Reference Standard RS 2x30, RS 1x30
User’s Guide
RS-UG91-EN www.appliedp.com
50
Commands for direct control of SU/SI contacts:
SYSTem:ENERgy:DOSage:CONTact {<SI>},{<SU>}
Direct setting of state (0=open / 1=closed) of SI and SU contacts.
SYSTem:ENERgy:DOSage:CONTact?
Request for state (0=open / 1=closed) of SI and SU contacts.
SYSTem:ENERgy:DOSage:CONTact:VOLTage {<SU>}
Direct setting of state (0=open / 1=closed) of SU contact.
SYSTem:ENERgy:DOSage:CONTact:VOLTage?
Request for state (0=open / 1=closed) of SU contact.
SYSTem:ENERgy:DOSage:CONTact:CURRent {<SI>}
Direct setting of state (0=open / 1=closed) of SI contact.
SYSTem:ENERgy:DOSage:CONTact:CURRent?
Request for state (0=open / 1=closed) of SI contact.
Commands for reading status of dosage function:
SYSTem:ENERgy:DOSage?
Request for general status of dosage function. If dosage function is active reply is: percentage, done
energy or time and full energy or time.
SYSTem:ENERgy:DOSage:STATus?
Request for status of dosage function. Reply is 0 (dosage not active), 1 (dosage initialized), 2 (dosage
started and running) or 3 (dosage stopped).
SYSTem:ENERgy:DOSage:STATus:PERCentage?
Request for done percentage of dosage.
SYSTem:ENERgy:DOSage:STATus:VALue?
Request for done value (time or energy amount) of dosage.
SYSTem:ENERgy:DOSage:STATus:TIMe?
Request for left time of time dosage in seconds.
SYSTem:ENERgy:DOSage:STATus:ENERgy?
Request for left energy of energy dosage in Ws resp. VAs.
SYSTem:ENERgy:DOSage:STATus:ENERgy:KWH?
Request for done energy of energy dosage in kWh resp. kVAh.