28
MEASure:CURRent:AC?
Request for the measured current value,
MEASure: CURRent:AC:PHASe?
Request for the phase of measured current,
MEASure: CURRent:AC:DISTortion?
Request for the distortion of measured current,
MEASure: CURRent:AC:SUM:SQuare?
Request for the sum of squares of measured currents.
MEASure:POWer:AC[:ACTive]?
Request for the measured active power value,
MEASure:POWer:AC:REACtive?
Request for the measured reactive power value,
MEASure:POWer:AC:APParent?
Request for the measured apparent power value,
MEASure:POWer:AC:FACTor?
Request for the power factor value,
MEASure:POWer:AC:SUM:ACT?
Request for the sum of measured active powers,
MEASure:POWer:AC:SUM:REACtive?
Request for the sum of measured reactive powers,
MEASure:POWer:AC:SUM:APParent?
Request for the sum of measured apparent powers.
MEASure:IMPedance:AC:MODule?
Request for the impedance module value,
MEASure:IMPedance:AC:PHASe?
Request for the impedance phase value,
MEASure:IMPedance:AC:REAL?
Request for the impedance real part value,
MEASure:IMPedance:AC:IMAGinar?
Request for the impedance imaginary part value.
MEASure:FREQuency?
Request for measured frequency.
MEASure:FREQuency:CHANnel?
Request for channel on which frequency was measured. Answer is channel number (0-5).
MEASure:FREQuency:MODe?
Request for frequency measuring mode. Answer is 0 (automatic) or 1 (fixed freq).
MEASure:FREQuency:MODe {<mode>}
Setting of frequency measuring mode. Parameter is 0 (automatic) or 1 (fixed freq).
MEASure:FREQuency:MODe:VALue?
Request for frequency in fixed frequency measuring mode. Answer is frequency in Hz.
MEASure:FREQuency:MODe:VALue {<freq>}
Setting of frequency in fixed frequency measuring mode. Parameter is frequency in Hz.
MEASure:FREQuency:MODe:CHANnel?
Request frequency channel mode. Answer is 0 (auto), 1 (voltage), 2 (current) or 3 (fixed).
MEASure:FREQuency:MODe:CHANnel {<mode>}
Setting of frequency channel mode. Parameter is 0 (auto), 1 (voltage), 2 (current) or 3 (fixed).
MEASure:FREQuency:MODe:CHANnel:FIX?
Request for channel in fixed frequency channel mode. Answer is channel number (0-5).
MEASure:FREQuency:MODe:CHANnel:FIX {<chnl>}
Setting of channel in fixed frequency channel mode. Parameter is channel number (0-5).

Table of Contents