7
Preparations
Actions Before First Use (6 steps)
1. Visually check the package for damage. If the
product looks damaged due to freight, contact
the transport agency.
Do NOT use the product.
2. The packaging should be recycled according to
local regulations.
3. Read this IFU.
4. Choose a designated area, where this IFU
should be kept and is easily accessible at all
times.
5. Verify each functions (Up-Down, Open-Close
lift legs)
6. Initiate the emergency lowering function to
check it is working correctly. To do this, pull up
the red emergency lowering handle and keep it
pulled up, whilst at the same time, push down
on the lifting arm until it starts lowering.
NOTE
Initial activation may require additional
amount of load on the lifting arm. It may be
helpful if two people perform this check, one
to activate the handle and the other to push
down the lift arm.
Actions Before Every Use (2 steps)
1. Visually inspect Sara Flex. If any part is
damaged - do NOT use the product.
2. Check the battery level.
In between Use
Clean and disinfect the product according to section
“Cleaning and Disinfection Instructions” on
page 27.
Sara Flex Directions
Left and right
Sara Flex has a left and right side. (See Fig. 1)
Transfer direction
The caregiver must be positioned behind Sara Flex
during transfer.
Left
Right
Transfer
direction
Fig. 1