Step 3: Mix control solution by gently
inverting bottle. Remove the cap from the
control solution bottle. While the
Apply Blood symbol and Code No.
(F-2, F-3 or F-4) are alternating, place
a drop of the control solution on the
end of the test strip. DO NOT put control
solution on top of the test strip. The test
strip will draw up the solution and the
meter will beep. The meter will begin to
count down.
Step 4: In 5 seconds the meter will
beep again and display the control
solution result.Compare the result to
the range printed on the test strip bottle.
Make sure the result is within the
acceptable range. If the result falls
within this range, the meter and test
strip are working correctly.