Troubleshooting Control Solution Test
If the control solution test result is out-of-range, check the following and repeat the test:
Check Action
Did you do the test in If not, do the test again. Insert a
control solution mode? test strip; press the Orange/Up Arrow
Do you see “CONT” on button for control solution mode.
the screen with the result? “CONT” should appear on screen.
Have the test strips and/or Make sure that test strips and control
control solution expired? solutions are not past expiration date.
This date is shown on bottle.
Were control solutions at room If not, retest with new bottle
temperature (72-82°F (22-28°C)) of control solution, or warm up/cool
when used? down to room temperature.
Does the code on the test strip If not, obtain a new bottle
bottle match the code on of test strips.
the meter screen?