ASD Document 600540 Rev. J 74 FieldSpec® 3 User Manual
www.asdi.com Appendix A Troubleshooting
• Check that the Ethernet cable is securely inserted into the spectrometer
and host computer. Check that the Ethernet LED is on.
• For an isolated network, check that the Ethernet cable is a cross-over
Ethernet cable.
• For an established network, check that the Ethernet cable is a standard
Ethernet cable.
• Check that the IP Address is in the same range or subnet as the ASD
spectrometer. The same subnet means that the first three octets of the IP
address (xxx.xxx.xxx.___) match the spectrometer and the computer.
• Do a ping test to make sure the spectrometer is responding.
» Open up a command window by select Start->Run
»Type cmd in the Run window.
» Select OK to open the window.
» For an Ethernet connection, type:
ping 10.1.1.11
» For a wireless connection, type:
ping 10.1.1.77
Figure A-1 Run window with cmd.