ASD Document 600540 Rev. J 74 FieldSpec® 3 User Manual
www.asdi.com Appendix A Troubleshooting
Check that the Ethernet cable is securely inserted into the spectrometer
and host computer. Check that the Ethernet LED is on.
For an isolated network, check that the Ethernet cable is a cross-over
Ethernet cable.
For an established network, check that the Ethernet cable is a standard
Ethernet cable.
Check that the IP Address is in the same range or subnet as the ASD
spectrometer. The same subnet means that the first three octets of the IP
address (xxx.xxx.xxx.___) match the spectrometer and the computer.
Do a ping test to make sure the spectrometer is responding.
» Open up a command window by select Start->Run
»Type cmd in the Run window.
» Select OK to open the window.
» For an Ethernet connection, type:
ping 10.1.1.11
» For a wireless connection, type:
ping 10.1.1.77
Figure A-1 Run window with cmd.