102
HI-Limit
This message appears on the display if the DUT measurement exceeds the HI-Limit
setting of any parameter (except AC withstand).
LO-Limit
This message appears on the display if the DUT measurement drops below the LO-
Limit setting of any parameter (except AC withstand).
HI-Lmt T
This message appears on the display if the DUT leakage current exceeds the HI-Limit
setting, of the Total current parameter in AC withstand.
LO-Lmt T
This message appears on the display if the DUT leakage current drops below the LO-
Limit setting, of the Total current parameter in AC withstand.
HI-Lmt R
This message appears on the display if the DUT leakage current exceeds the HI-Limit
setting, of the Real current parameter in AC withstand.
LO-Lmt R
This message appears on the display if the DUT leakage current drops below the LO-
Limit setting, of the Real current parameter in AC withstand.
CONT-Fail
This message appears on the display if the DUT fails the basic Continuity check
performed during an AC/DC Withstand test when Continuity is selected ON.
Arc-Fail
This message appears on the display if the DUT arcing current exceeds the Arc
Sense limit (Arc Sense = 1…9) and the Arc function is active.
Short
This message appears on the display if the DUT current is well beyond the metering
range of the test.
Charge-LO
This message appears on the display if the leakage current during Ramp-up falls
below the Charge -LO setting.
Breakdown
This message appears on the display if the DUT current is well beyond the metering
range of the test and the arcing condition beyond the arc sense limit.
Interlock Open
This message appears on the display if the Remote Interlock feature is activated
before or during a test. The Remote Interlock feature utilizes a set of closed contacts