100
5.4. Perform Tests Metering
Each test performed by the OMNIA contains a unique set of parameters and therefore
requires specialized metering for each test. The following table describes what
meters will be displayed for each of the different test types.
METERING SCREENS 8204/8206/8207
TEST TYPE
CURRENT(T)
CURRENT(R)
VOLTAGE
TIME
RESISTANCE
AC
Withstand
X
X
X
X
DC
Withstand
X
X
X
IR
X
X
X
Continuity
X
X
Ground
Bond
X
X
X
ADDITIONAL METERING SCREENS MODELS 8206/8207
TEST TYPE
VOLTAGE
CURRENT
POWER
TIME
PF
LEAKAGE
Run Test
X
X
X
X
X
X
ADDITIONAL METERING SCREENS MODELS 8206/8207
TEST TYPE
VOLTAGE
LEAKAGE
TIME
MD-VOLT
I-MAX
Leakage
Current
X
X
X
X
X
The meter layout on the Perform Tests screen can be customized to show the most
relevant information in the larger meter windows. Please refer to Section 4.
Programming Instructions for more information.
5.5. Displayed Messages
Directly above the Metering screens is the Test Status display. This portion of the
display is active during the test and allows you to view the type of test being
performed and status of the test step. At the end of a test, the Test Status display will
either inform you that the test has passed or give a direct indication of the kind of
failure that occurred during the test. The following is an example of the Test status
display:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8206 and is the answer not in the manual?

Associated Research OMNIA II 8206 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8206
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals