121
ADD2 GND,Current,Voltage,HI-Limit,LO-Limit,Dwell Time,Offset,Scanner
Channel,Frequency,DualCHECK
ADD2 GND,25.00,8.00,100,0,1.0,0,0,60,OFF
Adding a Continuity Test
ADD2 CONT,HI-Limit,LO-Limit,Dwell,Offset,Scanner Channel,Continuity
ADD2 CONT.,1000,0.00,1.0,0.00,OOOOOOOO
Adding a Functional Run Test
ADD2 RUN,Voltage-HI,Voltage-LO,Amp-HI,Amp-LO,Dwell,Delay,Leakage-
HI,Leakage-LO,Power-HI,Power-LO,PF-HI,PF-LO,Active Link,PLC
Control,Voltage,Frequency,A-Hi,Output N/G,Volt Rang,OC-Fold
ADD2 RUN,125.0,0.0,10.00,0.00,1.0,0.2,10.00,0.00,1000,0,1.000,0.000,OFF,M1
Adding a Leakage Current test
1,2
ADD2 LCT,Leakage-HI,Leakage-LO,Voltage-HI,Voltage-LO,Dwell Time,Offset,Delay
Time,Neutral,Reverse,Ground,Meas.Device,Probe,Peak/RMS,Active Link,AC/DC,
Ranging Mode,PLC Control,Scanner Setup
ADD2 LCT,6000,0.0,125.0,0.0,0.5,1.0,0.0,1.0,CLOSED,OFF,CLOSED,UL544NP,
Ground To Line,RMS,OFF,AC+DC,Auto,M1,OOOOOOOO
1
PLC Control is only available if option 04 (3 Remote Memory Send) or option 05 (7 Remote Memory
Send) is installed.
2
Please refer to the table above for additional optional parameters available with installed options.