EDOC
This command is used to control the Continuity and Ground Bond tests at the DUT
outputs of the LCT and RUN test interface. It should only be used when Continuity or
Ground Bond tests are routed through the HS-16 external, PC controlled scanner.
When the EDOC command is set to 0 the Continuity or Ground Bond test will be
disconnected from the GND and Case terminals of the DUT-outputs to eliminate
possible parallel paths with the scanner outputs.
7.4.5. System Parameter Editing Commands and Companion Queries
These commands are used to modify the system parameters for the instrument.
These commands require a parameter value to be included with the command. The
companion query command will read the parameter using the same value that is used
for setting the parameter.